Importing pulse programs from xwinnmr to Topspin 3.2

Had some learning experience while importing pulse programs written in xwinnmr to topspin 3.2. While applying hard pulses just after the shaped pulse on a channel, it is mandatory to specify the hard pulse power again before you apply the hard pulse. Or in other words, last point of shaped pulse line is 0W and so your hard pulse would also be of 0W if you do not specify hard pulse power after the shaped pulse line. Interesting to note that you would not get any error message in compilation of the pulse program, its just that your program will not work.

This problem was identified by an NMR expert Alex Hansen, so thought it would be best to keep the solution in public.